论文标题
自动导管尖端重新定位,用于心内超声心动图
Automated Catheter Tip Repositioning for Intra-cardiac Echocardiography
论文作者
论文摘要
目的:心内超声心动图(ICE)是指导心脏电生理学和结构心脏干预的强大成像方式。 ICE提供了对解剖和设备的实时观察,同时可以直接监测潜在并发症。在单一操作员的设置中,医师需要在冰导管和治疗设备之间进行来回切换,从而使连续的冰支撑不可能。因此,有时会实施两个操作员的设置,这是增加房间职业和成本的挑战。有时会实施两个操作员设置,但增加了程序成本和房间职业。 方法:冰导管机器人控制系统是通过自动导管尖端重新定位(即查看恢复)方法开发的,该方法可以再现先前导航到并保存的重要视图。在心脏幻像和动物实验的结合下,证明并评估了所提出的方法的性能。 结果:自动化冰视恢复达到的导管尖端位置精度为2.09 +/- 0.90 mm,动物研究中的3.93 +/- 2.07度的导管图像方向精度和0.67 +/- 0.79 mm和0.37 mm和0.37 +/- 0.37 +/- 0.19心脏幻象研究学位。我们所提出的方法在动物无并发症的情况下也成功地使用了,这表明使用冰导管机器人的无氟二骨穿刺的可能性。 结论:机器人冰成像有可能提供精确且可重复的解剖学视图,这可以减少总体执行时间,手术的劳动负担以及一系列心脏手术的X射线使用。关键字:自动化视图恢复,路径计划,心内超声心动图(ICE),导管,肌腱驱动的操纵器,心脏成像
Purpose: Intra-Cardiac Echocardiography (ICE) is a powerful imaging modality for guiding cardiac electrophysiology and structural heart interventions. ICE provides real-time observation of anatomy and devices, while enabling direct monitoring of potential complications. In single operator settings, the physician needs to switch back-and-forth between the ICE catheter and therapy device, making continuous ICE support impossible. Two operators setup are therefore sometimes implemented, with the challenge of increase room occupation and cost. Two operator setups are sometimes implemented, but increase procedural costs and room occupation. Methods: ICE catheter robotic control system is developed with automated catheter tip repositioning (i.e. view recovery) method, which can reproduce important views previously navigated to and saved by the user. The performance of the proposed method is demonstrated and evaluated in a combination of heart phantom and animal experiments. Results: Automated ICE view recovery achieved catheter tip position accuracy of 2.09 +/-0.90 mm and catheter image orientation accuracy of 3.93 +/- 2.07 degree in animal studies, and 0.67 +/- 0.79 mm and 0.37 +/- 0.19 degree in heart phantom studies, respectively. Our proposed method is also successfully used during transeptal puncture in animals without complications, showing the possibility for fluoro-less transeptal puncture with ICE catheter robot. Conclusion: Robotic ICE imaging has the potential to provide precise and reproducible anatomical views, which can reduce overall execution time, labor burden of procedures, and x-ray usage for a range of cardiac procedures. Keywords: Automated View Recovery, Path Planning, Intra-cardiac echocardiography (ICE), Catheter, Tendon-driven manipulator, Cardiac Imaging